Turn the unit on.
-
'SELECT A MODE' will be displayed.
-
Press the 2Continuous button: Unit will calibrate itself for temperature and incline, do
not move the unit during this time.
-
'+0.000 G' or '-0.000 G' will be displayed on the top line. Unit may now be moved. The
bottom line of the display shows the peak value only.
-
Rotate the unit 180° so its bottom is flat on the level
surface and the display is facing you.
-
'0.000 G' ±0.010
will be displayed on the top line. Calibration check step 1 is complete.
-
Rotate the VC2000 90° so the front panel is facing up and
parallel to the level surface. If there is a switch or connector on the rear panel the unit
will not set flat; Therefore, allow that portion of the unit to hang
off the edge of the level surface. Now only approximately ½ of the unit will be resting on the
level surface.
-
Press the CLEAR button: 'SELECT A MODE' will be displayed.
-
Press the 2Continuous button: Unit will calibrate itself for temperature and incline, do
not move the unit during this time.
-
'+0.000 G' or '-0.000 G' will be displayed on the top line. Unit may now be moved.
-
Rotate the VC2000 180° so the front panel is facing down
and parallel to the level surface, allowing the display portion of
the unit to hang over the edge of the level surface.
-
Using a mirror '+2.000 G' ±0.020 will be displayed on the
top line. Calibration
check is completed.