Frequently Asked Questions
Click here for Profile 3 FAQ
Q. Profile will not import data.
A. One or
more of the following:
- Download the
latest Profile software from the Profile Pro support page.
- USB serial
converter drivers not installed correctly. The
drivers are available on the Profile Support page of our web site or on the Profile CD.
- The VC4000 is not
turned on or is unplugged.
- Wrong COM Port
selected. Go to VC4000 tab|Communications Manager and select
Auto or the correct COM port, usually the last one in the list if
using USB.
- The USB port
drivers are not loaded yet. Wait about 5 seconds after plugging
the VC4000 in and turning it on before attempting to import.
- PDA
synchronization software such as "Hot Sync" is running and
controlling the serial port. Close the PDA software and try
again.
- Wrong cable or
not connected properly.
- Open Control
Panel/Device Manager and check COM ports. “VC4000 Virtual COM port (COMx)"
should be listed. If it is not, check other devices and reinstall
drivers if USB serial port has a question mark.
Q. I had a sensor set to zero adjust by mistake. How can I
correct this after the test is completed?
A. Go into the sensor settings for the run and make the minimum range
the opposite sign of the maximum range. For example, in the
case of a 300lb load cell, set the minimum range to -300 and maximum
range to 300. The resolution of the sensor results will now be
half of what it would have been.